Herunterladen Inhalt Inhalt Diese Seite drucken

Error Messages - sylvac Hi-Cal 300 Bedienungsanleitung

Höhenmessgerät
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

19 Error messages

Err0
Sensor measuring error. Press SET , and control the measuring reference
Err1
Parity Error (RS232 | USB connexion)
Err2
Unkown command (RS232 | USB connexion)
Err6
Error during the probe measuring detection sensor procedure. In case of
a problem during Hall initialization, please proceed with following
controls:
- Is the probe free to move ?
- Is the measuring force correctly set (see also chapter 17) ?
- If REF1 and CMDE are displayed, paddles may have been activated during
the initialization procedure. Please check that the paddles are free.
Err7
Possible action on the Keypad or FCT key during initialization. Please check
that keypad and FCT are free, then restart the Hi_Cal.
REF1
paddles may have been activated during the initialization procedure. Please
CMDE
check that the paddles are free, and restart the Hi_Cal.
53

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Hi-cal 150

Inhaltsverzeichnis