Herunterladen Inhalt Inhalt Diese Seite drucken

Interacoustics Callisto Gebrauchsanweisung Seite 54

Vorschau ausblenden Andere Handbücher für Callisto:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 69
Callisto™ Instructions for Use - EN
Medical CE-mark:
Real Ear Measurement Standards:
Stimuli:
Frequency range:
Frequency accuracy:
Distortion:
Intensity range:
Intensity accuracy:
Measurement Intensity Range:
Frequency Resolution:
Probe microphone:
Reference microphone:
Intensity Accuracy:
Cross talk
Available tests:
Compatible Software:
REM440 software - technical specifications
The CE-mark indicates that Interacoustics A/S meets the requirements of Annex II of
the Medical Device Directive 93/42/EEC. Approval of the quality system is made by
TÜV – identification no. 0123.
IEC 61669, ISO 12124, ANSI S3.46.
Warble Tone, Pure Tone, Random noise, Pseudo random noise, Band limited white
noise, Chirp, ICRA, Real Speech, any other sound file (automatic calibration available).
100Hz – 10kHz
Less than ± 1 %
Less than 2%
40 – 90 dB
Less than ± 1.5 %
Probe microphone 40-140 dB SPL ± 2 dB.
1/3, 1/6, 1/12, 1/24 octave or 1024 point FFT.
Intensity: 40 – 140 dB
Intensity: 40 – 100 dB
Less than ± 1.5 dB
Cross talk in the probe and probe tube will alter the obtained results with less than 1
dB at all frequencies.
REUR
REIG
RECD
REAR
REAG
Noah 4, Noah 3.7, OtoAccess
REOR
REOG
REUG
Input – Output
TM
and XML compatible