6. Commissioning, operation
3. Change the 4 characters using the ▲ arrow key.
4. Press the ▼ arrow key to go to the next position.
5. When all 4 characters are entered as required, confirm by pressing ENTER/MENU.
A small 2 appears and the following 4 characters can be changed as well.
6. Use ESC to return to the measuring mode.
When accessing oP1 using ENTER/MENU, the oP1 function (calibration by
code) is activated, even if the menu has been exited by pressing ESC.
Example of the measured value display after entering a probe calibration code:
The small 1 in the left corner, in combination with the display segment CAL in the centre
of the display, indicates that the calibration by code oP1 has been activated.
Physical calibration oP2
1. Use the ▲▼ arrow keys to select oP2.
2. Press ENTER/MENU to confirm the desired setting.
1 P is displayed in the lower part of the display.
3. Use the ▲▼ arrow keys to select from 1-point 1 P-, 2-point 2 P- and 3-point 3
P-calibration.
WIKA operating instructions, models CTH6300 and CTH6500
oP2
1 P
2 P
3 P