For further informations see PC software DME 4, menu bar
"Help".
To perform an output calibration you have to
connect a voltmeter respectively ammeter of
sufficient accuracy to the output terminals.
Current output / Voltage output
(Output A: x = 1, output B: x = 2, output C: x = 3, output
D: x = 4)
Variantes
Output
Brx01 Rx43
0 Ω
Output
(950685)
open
mA
or
soldered
sol-
Output V
open
dered
Calculation of resistors Rx45 and Rx46 for the full-scale
output currents Y2 in the range ≥ 1 to ≤ 20 mA:
Current output
Rx34 // Rx46 =
Y2 [mA]
0.99158 V
Y2
Rx45
Order No.
20
mA
open
10
mA
open
5
mA
open
2.5 mA
2.7 kΩ
951 245
1
mA
3.3 kΩ
951 253
Calculation of resistors Rx34 and Rx44 for the full-scale
output voltages Y2 in the range ≥ 1 to ≤ 10 V:
Voltage output
Rx34 // Rx44 = Y2 [V] · 27'229.4
Y2
Rx34
Order No.
10
V
270 kΩ
951 485
5
V
270 kΩ
951 485
2.5 V
68 kΩ
951 419
1
V
27 kΩ
951 360
38
Rx34
Rx44
Rx45
Rx46
27 kΩ
open
variable
variable
(951360)
0 Ω
(950685)
never
variable
variable
or
mind
soldered
1
1
–
27 kΩ
Rx46
Order No.
49.9 Ω
–––
102 575
Ω
–––
100
951 089
Ω
–––
200
101 717
Ω
470
951 162
1.5 kΩ
951 211
Rx44
Order No.
open
–––
270 kΩ
951 485
open
–––
open
–––
The locations of the variable components on the plug-in
board are shown in Figure 13.
R446
R445
R443
R343
Br301
Br201
Br401
R344
R444
R434
R334
D
C
Fig. 13. Location of the variable components.
7. Maintenance
No maintenance is required.
8. Dimensional drawing
M2.5 x 10
14
2
173.5
160
Fig. 14. EURAX DME 424/442, front plate width 14 TE.