PC board, terminal, housing plug
Exchange of housing plug to FHX40
3 mm
Unplug cable from FHX40.
Unscrew cover.
Disconnect cables from PC board and unplug display cable (FHX40).
Loosen the 3 screws at the PC board until it can be lifted up. 
It is not necessary to disconnect the cables from the bottom side of PC board.
If a impact protection is installed, loosen the screw and pull it off.
For the new housing plug it is not necessary to mount the impact protection agian.
Unscrew the housing plug.
Re-assembly is carried out in reverse order.
Please note:
Keep the display cable below the PC board.
Tighten the screws, holding the PC board, hand tight only!