Setting operating mode

RET basic safety control

Operating the device in mode A or B
Mode A
All settings will be stored if the device is switched off or discon-
nected from the power supply. The agitation and heating functions
will be set to OFF when the device is powered on.
Mode B
All settings will be stored if the device is switched off or discon-
nected from the power supply. The agitation and heating functions
will be set to ON or OFF when the device is powered on, depen-
ding on the previous status of the device.
Factory setting: mode A
The mode selected will be shown on the dis-
play when the device is started up.
Change the mode
☞ Move device switch (A) to the OFF position
☞ Press and hold rotary knobs (C and D)
☞ Move device switch (A) to the ON position
☞ Release rotating knobs (C and D)
➣ The set value is indicated on the display (F)
i
xx
REO basic
Operating the device in the preset mode
The operating mode is preset and not changeable.
Description of the mode and the display see RET
Factory setting REO basic: mode B
18
xx
A
(x: software version)
mode
A
Setting the safe temperature limit for
RET basic safety control
The safe temperature limit that has been set
will be displayed when the device is switched
on.
Factory setting: 360 °C
Adjustment range: > 50 - 360 °C
Setting the safe temperature limit
☞ Move device switch (A) to the ON position
☞ The safe temperature limit can be adjusted using a screwdriver
➣ The set value is indicated on the display (F)
i
150
SAFE
The temperature set for the heating plate (maximum 340 °C) will
be at least 10 °C lower than the safe temperature limit.
Warning: The safe temperature limit must always be set at
least 25 °C lower than the flash point of the media to be pro-
cessed.
Functional check of inactivating the safety circuit
☞ Position the safe temperature limit at the left (50 °C)
☞ Move device switch (A) to the ON position
➣ Indication on the display (F)
i
360
SAFE
Fig. 5
Err 25
092008