Herunterladen Inhalt Inhalt Diese Seite drucken

Seca 677 Bedienungsanleitung Und Garantieerklärung Seite 29

Vorschau ausblenden Andere Handbücher für 677:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 1
Evaluating Body Mass Index
Compare the value determined with the categories quoted below which correspond to
those used by WHO, 2000.
• BMI below 18.5:
Warning, this patient weighs too little. There could be a tendency to anorexia. An in-
crease in weight is recommended to improve well-being and performance. If in doubt,
consult a specialist.
• BMI between 18.5 and 24.9:
The patient is of a normal weight. He can stay as he his.
• BMI between 25 and 30 (pre-obese):
The patient is slightly to moderately overweight. He should cut his weight if he is al-
ready ill (e.g. diabetes, hypertension, gout, fat metabolism disorders).
• BMI over 30:
Weight reduction is essential. Metabolism, cardiovascular system and bones are all un-
der strain. A consistent diet, plenty of movement and behavioural training are all rec-
ommended. If in doubt, consult a specialist.
Weighing with pre-Tare
If the pre-Tare function is activated, a stored weight value is deducted from the currently
measured weight. This function is useful, for example, if the weight of a wheelchair is not
to be displayed.
– Press the Start key with no load on the
scale.
se(a
8. 8 . 8 . 8 . 8 .
,
secutively in the display. The scale is
then automatically set to zero and ready
for operation.
– Press the FUNC key.
The last function used appears in the
display.
– Use the arrow keys to select the pre-
Tare function, FU 3. Activate the func-
tion by pressing the FUNC key again.
In the display you will see the tare value
from the memory.
– You can use the arrow keys to adjust
the value in increments according to the
graduations displayed.
– Once the correct value is set, press the
FUNC key again. The pre-Tare function
is now activated.
– Move onto the scale in the wheelchair
associated with that memory. Read the
weighing result off the digital display.
– To return to normal weighing mode,
press the FUNC key.
Model 677
:
0. 0
and
appear con-
29
GB

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis