Start Logging
In Logging Mode, the PosiTector DPM L+ automatically records
datasets into memory at user-selected time intervals. Select Start
Logging from the PosiTector App or menu, then enter the desired
logging measurement interval (between 1 minute and 8 hours) and
sync interval, then select OK. All previously logged datasets are
deleted from the probe's memory.
View Live Data
To view current sensor readings, select View Live Data from the
PosiTector App menu.
NOTE:
The first time a probe is connected, the instrument will
display Live Data. Once datasets have been logged, logged
datasets will be displayed.
To exit Live Data view, select View Log Data from the PosiTector
App menu.
Set Measurement Interval
The Measurement Interval determines
how often a dataset is recorded into
memory. The probe is able to store up
to 10,000 datasets in memory.
Set Sync Interval
The Sync Interval determines how often
the datasets in memory are synced
through WiFi to PosiSoft.net.
Auto-Batch Creation Interval
The Auto-Batch Creation Interval auto-
matically separates datasets into batch-
es based on daily, weekly, or monthly
intervals. Auto-Batch creation only
occurs when WiFi is on.
When a new Log is started the
PosiTector DPM L+ immediately
records the first dataset and
Logging Mode