Herunterladen Inhalt Inhalt Diese Seite drucken

VOLTCRAFT ET-100 Bedienungsanleitung Seite 42

Isolationstester
Vorschau ausblenden Andere Handbücher für ET-100:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Acoustic continuity check
Range
<35 Ω permanent audio signal
Overload protection 600 V
Insulation test
Nominal voltage
at output
(0 to + 10 %)
125 V
250 V
500 V
1000 V
Overload protection >/= 30 V
Max. number of measurements with full batteries:
125 V
250 V
500 V
1000 V
Do not exceed the permitted max. input values. Never touch circuits or parts of circuits with a
possible voltage of more than 33 V/ACrms or 70 V/DC! Danger to life!
42
Resolution
0,01 Ω
Measuring
Resolution
range
0,125 – 4 MΩ
0,001 MΩ
>4 – 40 MΩ
0,01 MΩ
>40 – 400 MΩ
0,1 MΩ
>400 – 4000 MΩ
1 MΩ
0,250 – 4 MΩ
0,001 MΩ
>4 – 40 MΩ
0,01 MΩ
>40 – 400 MΩ
0,1 MΩ
>400 – 4000 MΩ
1 MΩ
0,500 – 4 MΩ
0,001 MΩ
>4 – 40 MΩ
0,01 MΩ
>40 – 400 MΩ
0,1 MΩ
>400 – 4000 MΩ
1 MΩ
1 – 4 MΩ
0,001 MΩ
>4 – 40 MΩ
0,01 MΩ
>40 – 400 MΩ
0,1 MΩ
>400 – 4000 MΩ
1 MΩ
x 500
x 500
x 500
x 500
Test voltage
5.8 V
Accuracy
Nominal
test current
±(2,0% + 10)
±(2,0% + 10)
1 mA
±(4,0% + 5)
(at 125 kΩ)
±(5,0% + 5)
±(2,0% + 10)
1 mA
±(2,0% + 10)
±(3,0% + 5)
(at 250 kΩ)
±(4,0% + 5)
±(2,0% + 10)
±(2,0% + 10)
1 mA
±(2,0% + 5)
(at 500 kΩ)
±(4,0% + 5)
±(3,0% + 10)
1 mA
±(2,0% + 10)
±(2,0% + 5)
(at 1 MΩ)
±(4,0% + 5)
Test current
200 mA (typical)
Short-circuit
current
≤1 mA
≤1 mA
≤1 mA
≤1 mA

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis