Please read these instructions carefully before beginning the assembly. Failure to understand and
follow assembly instructions may result in injury to technician and/or end user and may void the
warranty. If you have any questions call Sunrise Medical Technical Support.
Tools required
1.
Allen wrench, sizes: 2.5 mm,
3mm
2.
Pliers
1
3
F
DEALER/TECHNICIAN WARNING
Attention dealers and qualiÞ ed technicians, do not operate or service this device without Þ rst reading the owners man-
ual. If you do not understand the instructions and warnings in the owners manual please contact the Sunrise Medical
Technical Service Department before operating and/or servicing the Empulse device. Failure to do so may result in
damage and/or injury. Find more information and important warnings in the device owner's manual or at:
www.SunriseMedical.co.uk
3.
Torque wrench
4.
8mm socket
(6-point or 12-point)
2
F
4
5
H
See page 28 for contact information
or visit www.SunriseMedical.co.uk
R90
ACTUATOR REPLACEMENT
®
INSTRUCTION SHEET
1. To remove handle: (Fig. 1)
a. Hold the handle in place so
the spring does not pop up
once the screws are re-
moved.
b. Using a 3mm Allen wrench,
remove the four screws (A)
that hold the handle (C) and
the latch wire holder (B).
c. Remove handle (C) and
spring (D).
To complete steps 2-5, set the de-
vice in a horizontal position (leave
battery in, battery not shown).
2. Using a 3mm Allen wrench,
remove the front bolts (E) that
mount the latch block (F). Note
the position of the bolts for reas-
sembly (Fig. 2)
3. Remove the latch block (F). (Fig.
3)
Note: There is a spring in the
latch block. It must be positioned
correctly for reassembly.
4. Next, remove the actuator.
a. Back off the actuator bolt (K)
by hand or with a plier. (Fig.
4)
b. Disconnect the connector
(L) of the actuator harness.
Then, remove the actuator.
(Fig. 4)
5.
Next, remove the cover and
actuator hardware.
a. Remove the cover bolt (G)
and cover (H) by sliding
toward the wheel using a
2.5mm Allen wrench. (Fig. 5)
b. Remove the bottom actua-
tor bolt (I) using a 3mm Allen
wrench. Next, remove the
spacer (J) underneath the
I
actuator . (Fig. 5)
J