Problem/question
You feel an unpleasant sensation
from the electrodes.
The skin in the treatment area
becomes red.

10. Technical specifications

Belt size for waist sizes from
Weight (belt and unit)
Electrode size
Parameters (500 Ohm load)
Pulse duration
Wave form
Power supply
Operating conditions
Storage
We reserve the right to make any technical alterations that are necessary in order to improve and develop the
product further. If the unit is used outside of the specifications, proper function is no longer guaranteed!
This unit complies with European standards EN60601-1, EN60601-1-2 and is subject to special precautions
regarding electromagnetic compatibility. Note that portable and mobile RF communication equipment can affect
this unit. For detailed information, contact customer service at the address provided.
The unit conforms to the requirements of the European Directive for Medical Products 93/42/EC, the MPG (Ger-
man Medical Devices Act). In accordance with the MPBetreibV (German Medical Device Operator Ordinance),
regular metrology checks must be performed if the unit is used for commercial or business purposes. Even for
private use, we recommend having the manufacturer perform a metrology check every 2 years.
Possible cause/remedy
The electrodes are poorly moistened. This can lead to skin irritation, as
the current is no longer distributed evenly over the entire surface. Switch
off the unit. Take off the belt, moisten the belt (electrodes) and your lower
back and restart the program according to the manual.
Discontinue the treatment immediately and wait until the skin condi-
tion has normalised. Temporary skin reddening under the electrode is
harmless and can be explained by the localised increase in blood flow.
However, if the skin irritation persists and you feel itching or irritation,
consult your doctor before continuing use. The cause may be an allergy
to the surface of the electrode.
Approx. 75 to 140 cm
Approx. 415 g with batteries
Approx. 81 x 65 mm
Output voltage:
Output current:
Output frequency: 2 – 110 Hz
60 – 220 μs per phase
Symmetric, biphasic rectangular pulses
4.5 V (3 x 1.5 V AAA, type LR 03)
0 °C to 40 °C, 20 to 65% relative humidity
0 °C to 55 °C, 10 to 90% relative humidity
18
max. 50 Vpp / 5.5 Vrms
max. 100 mApp / 11 mArms