CALIBRATION
The ACE KETOSCAN Lite must be calibrated after every
300 tests. After the 300th test, „Sen" is shown on the
display of the ACE KETOSCAN Lite and this means that
calibration or sensor replacement is necessary. After
the 300th test, the device is also locked because the
accuracy of the integrated sensor is then no longer
guaranteed. With a calibration, the 1-year warranty on
the sensor remains valid. You can send the device for
calibration to the service address given at www.keto-
scan.de.
We recommend sending it as a parcel, so that in cases
of loss an inquiry or proof of delivery is possible.
SENSOR REPLACEMENT SERVICE
The KETOSCAN Lite has a sensor that can be used up to
300 times to obtain accurate and stable measurement
results. The sensor replacement service, where the sen-
sor is completely recalibrated after replacement, is a
special care solution. It allows you to use the device
like a new one for a long time and avoid replacing the
entire device. Please contact the nearest service centre
or the shop where you purchased the machine to order
or have the service performed.
BATTERY NOTICES
The device associated with this operating manual con-
tains 2 AA batteries (1.5 V, Alcaline) in delivery condition.
These batteries are not made for recharging. The use
of a rechargeable battery is not possible with this de-
vice and can damage it. If the included batteries are
depleted, they must not be thrown into the household
waste. Used batteries may contain harmful substances
that can cause environmental or health damage. Plea-
se return all depleted batteries to electrical appliance
stores, who collect them for free.
49