DISPLAY INDICATIONS

DISPLAY
MEANING
bLo
Start to blow: Blow into the mouthpiece (or
drinking straw) until the unit vibrated.
SdC
Self-Diagnostic test Completed: Wait a few
seconds for the main test.
SEn
Sensor replacement required (operation
blocked): The unit has been used for 300
tests, now the sensor has to be replaced.
The service is required to get an accurate
result. Please contact the seller to order the
replacement. For this service you can send
the device to ACE Instruments (address see
www.ketoscan.de).
bAt
Low Battery Warning: Replace the 2 AA Alkali-
ne batteries against 2 new ones.
Er.2
Sensor-Error: The sensor is defective and has
to be exchanged. For this service you can
send the device to ACE Instruments
(see www.ketoscan.de).