Herunterladen Inhalt Inhalt Diese Seite drucken

Appropriate Usage; Preparation For Operation; Lc-Display - Unitest SLM MINI Bedienungsanleitung

Schallpegelmessgerät 93565
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Insert batteries supplied by respecting correct polarity. Batteries
with reversed polarity can lead to instrument destruction. Fur-
thermore, they may explode or ignite.
Avoid any heating up of the instrument by direct sunlight to en-
sure perfect functioning and long instrument life.
3.1

Appropriate Usage

The instrument may only be used under those conditions and for
those purposes for which it was conceived. For this reason, in
particular the safety references, the technical data including en-
vironmental conditions and the usage in dry environments must
be followed.
When modifying or changing the instrument, the operational
safety is no longer ensured.
4.0

Preparation for Operation

1.) Windbreak
The windbreak supplied should always be installed for outdoor
measurements, in case of strong air movements, or in dusty en-
vironments. Wind noise at the microphone results in erroneous
measurements. Furthermore, the windbreak protects the micro-
phone from humidity and mechanical shock.

2.) LC-Display

3.) Key On/ Off , MAX HOLD
Turn the unit off by holding down the key for approx 3 sec.
4.) Capacitor microphone
5.) Sliding switch MEAS/CAL measurement or internal calibrator
6.) Adjustment potentiometer
7.) Battery Cover
4.1
LC-Display
8.) Battery capacity display
9.) MAX: measurement data display of maximum value
10.) dBA: unit
11.) Measurement data display of current sound level
12.) OVER: exceeding the measurement range > 130 dB
13.) FAST: time evaluation FAST
14.) UNDER Under-running the measurement range < 32 dB
6

Werbung

Kapitel

Inhaltsverzeichnis
loading

Inhaltsverzeichnis