Herunterladen Inhalt Inhalt Diese Seite drucken

GERATHERM cardio connect Gebrauchsanweisung Seite 58

Inhaltsverzeichnis
Verfügbare Sprachen

Verfügbare Sprachen

  • DEUTSCH, seite 5
eMc GuIDance
table 1
Guidance and manufacturer's declaration - electromagnetic emissions
emissions test
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions IEC 61000-3-2
Voltage fluctuations / flicker
emissions IEC 61000-3-3
table 2
Guidance and manufacturer's declaration – electromagnetic Immunity
Immunity test
Electrostatic discharge
(ESD) IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
EN
Surge IEC61000-4-5
Voltage dips, short in-
terruptions and voltage
variations on power
supply input lines
IEC 61000-4-11
compliance
Group 1
Class [ B ]
Class A
Comply
Iec 60601-1-2
test level
±8 kV contact
±2 kV, ±4 kV, ±8 kV,
±15 kV Luft
±2 kV for power supply
lines
±1 kV signal input/
output
100 kHz repetition
frequency
±0,5 kV, ±1 kV
differential mode
±0,5 kV, ±1 kV, ±2 kV
common mode
0 % UT; 0,5 cycle.
At 0°, 45°, 90°, 135°,
180°, 225°, 270° and
315°.
0 % UT; 1 cycle and
70 % UT; 25/30 cycles;
Single phase: at 0°.
0 % UT; 250 / 300
cycle
52
compliance level
±8 kV contact
±2 kV, ±4 kV, ±8 kV,
±15 kV Luft
±2 kV for power supply
lines
Not applicable
100 kHz repetition
frequency
±0,5 kV, ±1 kV
differential mode
0 % UT; 0,5 cycle.
At 0°, 45°, 90°, 135°,
180°, 225°, 270° and
315°.
0% UT; 1 cycle and
70% UT; 25/30 cycles;
Single phase: at 0°.
0 % UT; 250 / 300
cycle