Guidance and manufacturer's declaration - electromagnetic immunity

The FlexScan L362T is intended for use in the electromagnetic environment specified below.
The customer or the user of the FlexScan L362T should assure that it is used in such an environment.
Immunity test
Electrostatic
discharge (ESD)
EN61000-4-2
Electrical fast
transient / burst
EN61000-4-4
Surge
EN61000-4-5
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
EN61000-4-11
Power frequency
(50/60Hz)
magnetic field
EN61000-4-8
NOTE U
is the a.c. mains voltage prior to application of the test level.
T
IEC60601 test level
±6kV contact
±8kV air
±2kV
for power supply lines
±1kV
for input/output lines
±1kV line(s) to line(s)
±2kV line(s) to earth
<5% U
(>95% dip in
T
U
) for 0.5 cycle
T
40% U
(60% dip in
T
U
) for 5 cycles
T
70% U
(30% dip in
T
U
) for 25 cycles
T
<5% U
(>95% dip in
T
U
) for 5sec
T
3A/m
Compliance level
±6kV contact
±8kV air
±2kV
for power supply lines
±1kV
for input/output lines
±1kV line(s) to line(s)
±2kV line(s) to earth
<5% U
(>95% dip in
T
U
) for 0.5 cycle
T
40% U
(60% dip in
T
U
) for 5 cycles
T
70% U
(30% dip in
T
U
) for 25 cycles
T
<5% U
(>95% dip in
T
U
) for 5sec
T
3A/m
Electromagnetic environment -
guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the
FlexScan L362T requires continued
operation during power mains
interruptions, it is recommended
that the FlexScan L362T be
powered from an uninterruptible
power supply or a battery.
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a
typical commercial or hospital
environment.