36
• Ensure that the medium is well mixed (high speeds and large
• The temperature sensors (PT 100, PT 1000) should be immersed
• When using metal vessels, do not place the temperature sen-
an audible signal will be given
• With highly viscous reaction media, the distance between the
14
Tips
stirring bars ensure good mixing). A medium which is well mixed
will ensure uniform temperature distribution and better tempe-
rature control.
to the maximum possible depth. The minimum immersion depth
is 20 mm; do not go below this immersion depth.
sors on the bottom of the vessel. Placing sensors on the vessel
bottom can cause excessively high temperatures to be measu-
red especially in media which have poor conductivity. The tip of
the measuring sensor must be at least 5 mm from the vessel
bottom; a distance of 10 mm is ideal.
heating plate and the stirring bars must be kept as small as pos-
sible (to reduce the risk of stirring bar breakaway).