Important!
In case of a fault in the mobility of the
seat shell, always check the pelvis
mechanism first and then the foot rest
mechanism!
7.2.
Options
7.2.1. Lower extremities
Foot rest system
The foot rest system of the AKTIVLINE
tomized to the lower leg length of the patient� During
adjustment it must be ensured that the patient's feet are
comfortably positioned on the foot rest� The foot rest
system of the AKTIVLINE
as well as single-piece design, a split version must al-
ways be selected for a difference in leg length of 40 mm�
In order to adjust the foot rest system to the lower leg
length of the patient, release the screw on the clamp
block of the foot rest and shift the foot rest up or down,
as necessary, on the centered tubing of the foot rest
system; this also applies to the foot rest system with
individual foot supports� Then tighten the screw after
the adjustment is complete�
Calf support
Calf supports are already fitted ex-works�
Foot rest retraction unit
The foot rest retraction unit is located at the foot rest
suspension and mounted under the seating unit� The
stretching movement of the legs is restricted by a gas-
pressure spring, i�e� the gas-pressure spring restrains
stretching of the legs through resistance�
If you wish to release the foot rest retraction unit to al-
low free movement for the patient's legs, then proceed
as follows:
The mounting of the foot rest retraction unit is fit-
ted to the foot rest tube�
Release the quick pin by pressing the red button
and keeping it depressed� During pressing, pull the
pin from the opening�
This now hangs loosely attached to a suspension�
If you wish to refit the foot rest retraction unit, hold
the gas-pressure spring into the eyelet, which is
attached to the middle tube of the foot rest, with
one hand�
Then press the red button of the quick pin and
guide this into the opening/mounting intended for
this purpose�
52
ASSEMBLY / ADJUSTMENT / OPERATION
must be cus-
®
is available in both individual
®
Pull slightly to check whether the quick pin has en-
gaged and whether stretching of the legs is again
restrained by the gas-pressure spring�
Important!
Make sure that you do not stand or knee
directly in front of the foot rest during
assembly� Sudden stretching movements by the
patient could result in your being kicked or injured
unexpectedly�
Foot rest mechanism locking option
Activate the foot rest mechanism (mobility in the leg
area) by turning the clamp lever on the outside of the
seat shell counterclockwise until the foot rest can be
moved� Make sure not to turn the clamp lever more
than necessary, otherwise there is a risk of the lever
popping out of the thread� To deactivate the foot rest
mechanism, turn the clamp lever clockwise until the
foot rest is firmly locked again�
Important!
In case of a fault in the mobility of the
seat shell, always check the pelvis
mechanism first and then the foot rest
mechanism!
Option of adjusting foot angle
The foot angle adjustment is located on the connection
between the foot rest and the center tube of the foot
rest� You can adjust this as follows:
Turn the clamp lever counterclockwise to loosen it�
Now move the foot rest into the desired position�
Hold this position with one hand�
Now turn the clamp lever clockwise and tighten
firmly�