Herunterladen Diese Seite drucken

Siemens SITRANS F 7ME3011 Betriebsanleitung Seite 19

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

3
Connecting the Sensor Cable to the Transmitter
When connecting the sensor cable to the transmitter, make sure that the production numbers on the rating
plate of the transmitter are identical to those of the rating plate of the measuring pipe.
The transmitter, sensor cable and measuring pipe form one unit per flowmeter and have been calibrated
as such. If the measuring pipe and transmitter are not paired together correctly, a deviation in the
measured value may occur which is outside the limits of the calibration certificate. This does not affect the
principle function of the device.
.
The sensor cable is fixed perma-
nently to the measuring pipe. It is
pre--assembled ready to use and
may not be shortened or modified.
.
Excess lengths of cable must be
wound up taking into account the
minimum radius of 10 cm.
Figure 2
Connecting the sensor cable to the transmitter
SITRANS F - - Remote version
C79000-B5674-C115-02