... impossible to perform calibration.
... very slow response.
Dirt on the diaphragm or the mem-
brane.
De-hydrated membrane.
Electrode ageing or wear.
Clean, regenerate, replace the elec-
trode.
Storage
The connector should be protected from
humidity by the protective cap
Place the electrode in the storage protector
with KCl 3M.
Life of an electrode
The average life expectancy of a pH process
electrode oscillates between 6 or 12 months,
depending on the working conditions.
Electrodes exposed to high temperatures do
not last as long. Very alkaline mediums also
reduces the life span of electrodes.
If an electrode has too short life span,
less than three months, ask us about it.
Probably there is a more suitable elec-
trode for the application.
Specifications
Measuring range
Operating temp.
Max. pressure
Reference element
Diaphragm
Electrolyte
Body material
Thread
Connector
Min. immersion
depth
Applications
Waste water, galvanising industry, osmosis
processes, demineralized water, dirty or
viscous samples , etc.
CERTIFICATE OF QUALITY
The attached electrode 53 30 has passed
the quality test and follows the next speci-
fications:
• Asymmetry potential < ± 15 mV.
• Sensitivity, pH 4 to 7 (at 25 ˚C) > 98%.
• Response time, pH 4 to 7 < 20 s.
.
0 to 14 pH
0 to 100 °C
5 bar
cartridge
ring of porous PTFE
gel
glass
PG 13,5
S8
12 mm
GB
5