7.
RECEIPT OF CENTRIFUGES TO REPAIR
Health risk from contaminated equipment, rotors and accessories.
In case of returning the centrifuge for repairing to the manufacturer, please notice the following:
The centrifuge must be decontaminated and cleaned before the shipment for the protection of persons,
environment and material.
Decontamination certificate at goods return delivery (See APPENDIX)
We reserve the right to not accept contaminated centrifuges.
Further on all costs occurred for the cleaning and disinfection of the units will go to the debit of the customer's
account.
8.
TRANSPORT, STORAGE AND DISPOSAL
8.1
Transport
Before transporting, take out the rotor.
Only transport the unit in the original packaging.
Use a transport aid for transporting over longer distances to fix the motor shaft.
General transportation
8.2
Storage
During storage of the centrifuge the following environmental conditions must be observed:
in transport packaging
8.3
Transporting, Installing, Transferring and Disposing of the Centrifuge FC5515R
These instructions complement the previous general instructions in chapter 8 and do not replace them.
8.3.1
Transport
Please transport the device in the original packaging.
The centrifuge should always be transported by two people.
Use a transport aid for transporting over longer distances.
8.3.2
Installation
The weight of the centrifuge is about 35 kg. The centrifuge must always be transported by two people.
 Opening the carton and lifting out the device.
1. Cut the adhesive tape.
2. Open all 4 flaps of the carton.
3. Remove the accessories.
4. Reach with your hands under the device and lift the centrifuge from
the carton with another person.
 Place the device on a stable, horizontal and non-resonant lab bench.
1. Remove the front and back transport protection material.
2. Remove the plastic sleeve.
3. Observe a minimum distance of 30 cm to adjoining devices at the
sides and from the rear side to the wall.
4. Install the device in a well-ventilated location which is protected from
direct sunlight to prevent it from overheating.
Air temperature
rel. humidity
-25 to 60 ° C
10 to 75 %
Air temperature
rel. Humidity
-25 to 55 °C
10 to 75 %
Air pressure
30 to 106 kPa
Air pressure
70 to 106 kPa
EN-23