Maintenance
(a) In the case of pH sensors and manual temperature compensation, set the measurement
temperature.
(b) Enter the pH value or mV value of the buffer solution.
(c) Start calibration.
(d) The value is accepted once it has stabilized.
3.
Rinse the sensor with distilled water. Do not dry the sensor!
Calibration finished for ORP measurement.
pH only:
4.
Immerse the pH sensor into the second buffer solution (e.g. pH 4).
5.
Perform the calibration at the measuring device:
(a) Enter the pH value of the second buffer solution.
(b) Start the calibration.
(c) The value is accepted once it has stabilized.
The device calculates the zero point and slope and displays the values. Once the adjustment
values have been accepted, the device is adjusted to the new pH sensor.
6.
Rinse the sensor with distilled water.
7
Maintenance
7.1
Cleaning the sensor
Clean the sensor before each calibration.
Clean away fouling on the sensor as follows depending on the type of fouling:
1.
Oily and greasy films:
Clean with grease remover, e.g. alcohol, as well as hot water and (alkaline) agents
containing surfactants (e.g. dishwashing detergent).