Installation conditions
• Before screwing in the electrode, make sure the assembly thread is clean and runs
smoothly.
• Screw in the electrode until it is hand-tight (3 Nm). (Data apply only if installing in Endress
+Hauser assemblies.)
• Pay attention to the installation instructions provided in the Operating Instructions of the
assembly used.
Special instructions for the BP version
Before commissioning the electrode, you must remove the silicone seal from the junction.
 The system can only measure the pH value properly when the seal has been removed.
CAUTION
L
Glass electrode with pressurized reference
Possibility of sudden rupture and injury from glass splinters!
Always wear protective goggles when working with these electrodes.
Be particularly careful when removing the silicone seal on the reference junction. Here, a
knife is used to activate the electrode for measuring operation.
4.2
Installation angle
4.2.1
CPS16D and CPS96D
Do not install the electrodes upside down!
NOTICE
The angle of inclination must be at least 15° from the horizontal.
A smaller inclination angle is not permitted as it could cause an air bubble to form in the glass
sphere and prevent the inner electrolyte from completely wetting the pH diaphragm.
Select an installation angle for the electrode that is not smaller than 15°.
1.
Before screwing in the electrode, make sure the assembly threaded connection is clean
and runs smoothly.
2.
Screw in the electrode until it is hand-tight (3 Nm). (Data apply only if installing with
Endress+Hauser assemblies.)
3.
Also make sure to follow the installation instructions in the Operating Instructions of
the assembly used.