Herunterladen Diese Seite drucken

Hach AS950 Grundlegende Installation Und Wartung Seite 14

Vorschau ausblenden Andere Handbücher für AS950:

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 28
5.3 Plumb the sampler
Install the intake tube in the middle of the sample stream (not near the surface or bottom) to make
sure that a representative sample is collected.
1. For a sampler with the standard liquid detector, connect the tubing to the sampler as shown in
Figure
7.
Note: When Teflon-lined tubing is used, use the tubing connection kit for Teflon-lined PE tubing.
2. For a sampler with the optional non-contacting liquid detector, connect the tubing to the sampler
as shown in
Figure
Note: When Teflon-lined tubing is used, use the tubing connection kit for Teflon-lined PE tubing.
3. Install the intake tube and strainer in the main stream of the sample source where the water is
turbulent and well-mixed. Refer to
• Make the intake tube as short as possible. Refer to
intake tubing length.
• Keep the intake tube at a maximum vertical slope so that the tube drains completely between
samples.
Note: If a vertical slope is not possible or if the tube is pressurized, disable the liquid detector. Calibrate the
sample volume manually.
• Make sure that the intake tube is not pinched.
Figure 7 Plumbing—Standard liquid detector
14 English
8.
Figure
9.
Specifications
on page 3 for the minimum

Werbung

loading