To replace the power supply for the refrigerated sampler, refer to
Figure 16 Power supply replacement
7.6 Disposal
Child entrapment hazard. Remove the doors on the refrigerated cabinet before disposal.
Chemical exposure hazard. Dispose of chemicals and wastes in accordance with local,
regional and national regulations.
Section 8 Troubleshooting
8.1 General troubleshooting
Table 2
shows causes and corrective actions for several common problems.
Problem
No instrument power
26 English
D A N G E R
C A U T I O N
Table 2 Troubleshooting table
Possible cause
Problem with the main power
source.
Defective power supply
(refrigerated sampler only)
Defective controller
Figure
16.
Solution
Make sure that AC power gets to
the electrical outlet.
Replace the power supply.
Contact technical support.