DISPLAY INDICATIONS
SIGN
MEANING
Breath sampling error
This message is displaying lack of respiratory
sample for the analysis. The device will re­
enter to the measuring mode if you wait for a
moment. Blow a breath through the mouthpiece
for abput 3 seconds.
Low Battery warning
Charge the device using the provided USB
cable.
Battery charging (Blink)
A message displaying when the charge is
ongoing.
Complete charging the device
A message displaying when the charge is
completed.
Calibration Service is necessary
(Sensor Replacement Alarm)
A message alerts a calibration service is ne­
cessary. A calibration to maintain the precision
of the product is necessary, so, please contact
with a service center near your location at the
"Service Center" menu in the KETOSCAN app
or the purchasing store.
Sensor Error (E2)
An error message displaying when an error on
the sensor in the device is occurred. Please
contact with a service center near your loca­
tion at the "Service Center" menu in the KETO­
SCAN app or the purchasing store.
Bluetooth
mode
®
The device is connected with KETOSCAN app.