montHly cleanIng

1. pressurize the scuBA equipment.
2. If a full face mask is used, remove the cover by unscrewing the two screws. grasp
the lower part of the full face mask (where the valve is located), with the visor
upwards. turn the breathing valve clockwise to undo the bayonet coupling of the
breathing valve.
3. wash the full face mask and breathing valve, using a brush if necessary. use the
purge button to blow air through the valve to remove water and dirt. repeat until
the valve is clean and dry. for cleaning, Interspiro recommends using spirorinse
(336 921 322). other suitable cleaning agents may also be used. follow the
cleaning procedures as recommended by the manufacturer/supplier of the
cleaning agent.
Warning! Blow-off noIse And pressure surges mAy cAuse heArIng dAmAge. AlwAys use eAr
protectors when fIllIng And emptyIng AIr cylInders.
4. depressurize, disconnect and dry the full face mask and breathing valve.
cHeckIng tHe tIgHtness of tHe non-retUrn ValVe
the tightness of the non-return valve must be checked during monthly cleaning.
1. connect the breathing valve to the medium pressure supply hose of the regulator.
2. remove the breathing valve from the full face mask, or alternatively take
the mouthpiece of the breathing valve (see the "replacing the mouthpiece"
instructions in the "repair" section).
3. open the cylinder valve.
4. fill the free space of the non-return valve disc with water. (figure 15)
5. check that no water leaks into the breathing valve. the water level over the non-
return valve must not drop. If there is any leakage, clean or replace the non-return
valve and redo the check.
6. Blow the valve dry by pressing the purge button. check that the non-return valve
disc lies flat after the valve has been blown dry.
7. Close the cylinder valve and fit the breathing valve to a full face mask or
mouthpiece.
dIsInfectIng tHe fUll face mask/breatHIng ValVe
the full face mask/breathing valve must be disinfected during monthly cleaning or
when the need arises.
Interspiro recommends that full face masks and breathing valves used by more than
one person be cleaned and disinfected after each use.