Herunterladen Inhalt Inhalt Diese Seite drucken

Medela Symphony Bedienungsanleitung Seite 116

Vorschau ausblenden Andere Handbücher für Symphony:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 19
13. Appendix (cont.)
Electromagnetic Compatibility (EMC, IEC 60601-1-2:2007, Table 6)
116
Recommended separation distances between portable and mobile RF communications equipment
and the Symphony electric breast pump
The Symphony breast pump is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the Symphony electric breast pump can help
prevent electromagnetic interference by maintaining a minimum distance between portable and mobile
RF communications equipment (transmitters) and the Symphony electric breast pump as recommended
below, according to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter W
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in metres (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
Separation distance according to frequency of transmitter M
150 kHz to 80 MHz
d = 0.35√P
0.04
0.11
0.35
1.11
3.50
80 MHz to
800 MHz
d = 1.2√P
0.12
0.37
1.2
3.7
12
800 MHz to 2.5 GHz
d = 2.3√P
0.23
0.74
2.3
7.4
23

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis