• Press the Start measurement button
•
•
•
•
and in mm-Westergren.
•
• The device will scan the test Sedivette again and calculate the second measured value, which in turn is displayed both
in millimetres and in mm-Westergren.
Figure 44: Test measurement
To interpret the results, please compare the values displayed in your device with the following table:
Table 2
72-68 mm
Measured value 1:
40 mm - 44 mm
Measured value 1:
56 mm Westergren - 61 mm Westergren
Measured value 2:
63 mm - > 65 mm
Measured value 2:
109 mm Westergren - > 116 mm Westergren
Maintenance at the manufacturer.
• Measurement results must always be checked for plausibility by the operating personnel,
the laboratory management or the attending doctor.
• The general regulatory guidelines and statutory regulations for properly determining blood
sedimentation must be noted and complied with.
EN
100