Getting Started
11
Turn on the compressor and allow to pressurize, (see page 21.)
12.
Make sure your compressor is set to deliver the correct working pressure. The air pressure
regulator gauge should read 80 psi (5.5Bar). If adjustment is required, please see page 27
for compressor pressure adjustment instructions. The pressure is set at the iM3 factory.
13.
Place the foot control pedal on the floor in front of the unit. The foot pedal operates the High
and Low speed handpieces when depressed.
14.
Check handpiece operation:
a) Select HS on the Toggle, Remove High Speed Handpiece from holder and depress
handpiece foot control pedal—handpiece should now operate. Page 4 (C)
b) Select LS and Remove Low Speed Handpiece from holder and depress handpiece foot
control pedal—handpiece should now operate. Page 4 (C)
15.
Adjust individual handpiece pressure if required, (see page 16). Each handpiece—Low
speed and high speed has a specific pressure requirement.
16.
Pressurize water system by selecting the coolant bottle pressure switch to ON, located on
the front of the control panel. (see page 4 F)
Important. When filling distilled water and/or CLS Enviro bottles de-pressurize the
water system, by switching the coolant bottle pressure switch to off. Check water and air
supply to the 3 way air water syringe. Depress the right hand button for air and left hand
button for water or both together for a mist.
17.
Water volume. Remove high speed handpiece, from holder and depress foot control pedal.
Adjust water spray volume to a fine mist by rotating the flow control knob. (see page 4 "H")
Please observe manufacturers handpiece maintenance instructions for
handpiece longevity.