Handpiece Trouble Shooting

Low Speed Handpiece

PROBLEM
CAUSE
Handpiece
1. No or low air pressure.
won't run
2. Air line improperly sealed.
3. Inferior oil; too much or excess
more commonly too little oil or
water in the turbine.
4. Broken drive head.
Note new handpieces require a
"Run in time" this can be up to 20
hours of use
Low Speed Handpiece - 4:1 Straight Nose Cone
PROBLEM
Cant fit R&R pro-
phy angle
iM3 Inc
12414 NE 95th Street
Vancouver WA 98682 USA
Tel (360) 254 2981 Fax (360) 254 2940
CAUSE
1. Plastic drive shaft broken off
internally in Chuck housing.
2. Improper lubrication.
Only fit plastic iM3 R&R prophy
angles Code L7465.
www.im3vet.com
SOLUTION
1. Check the air line to be sure
the air is turned on and the
line is unobstructed.
2. Reconnect air line to
handpiece making sure the
gasket is fully contracted.
3. See lubrication instructions
on page 17.
4. Contact your local iM3
official Service Centre.
If the motor stops spinning
place a fine screw driver down
the nose and rotate shaft.
SOLUTION
1. Contact your local official
iM3 Service Centre.
2. See lubrication instruc-
tions, on page 17.