INFLATION
TOO LOW
MEASUREMENT
DELAYED
NO ECG
DETECTED
REPEAT BP
SERVICE
REqUIRED
SYSTEM ERRORS: A System Error (monitor displays a System Error screen with a
5-digit error code) indicates a possible problem with the monitor's hardware or
software. When this error occurs, the alarm will sound continuously. This alarm can
be muted by pressing the Down key. Pressing the Select key will clear the alarm, and
the monitor will rerun its internal tests. If the error occurs repeatedly, please contact
the SunTech Service Department or an authorized service agent.
To test the System Error alarm function, press the Up key when power is applied to
the monitor. System Error "30000" should appear on the screen. Pressing the Down
key will mute the alarm. Pressing the Select key will end the test; the monitor will
complete its power-up sequence and display the Main Measurement View.
K-sounds were
detected within
10mmHg of
the target
cuff inflation
pressure. BP
reported.
Due to safety
reason, a BP
measurement
could not be
taken at that
time.
The monitor is
not receiving an
ECG signal.
The BP
measurement
needs to be
repeated.
N/A
BP reading may be inaccurate. Advise the
patient to drop arm to the side during the
BP reading (avoid excessive bending). Take
another BP reading.
Note – Cuff will inflate to max pressure on
next BP reading.
The next BP reading will occur as scheduled.
• Review instructions for your stress system
interface with the Tango+ (See Tango+
Setup, page 7). Check to see that your
Stress System (if Custom, check the ECG
Trigger) is correctly set.
• Check that the cables referred to in the
instructions are correctly seated to their
connectors and show no signs of damage.
• Set the Waveform Display (under View
in the Main Menu) to ECG to verify the
monitor is receiving an ECG Signal.
Initiate a new BP reading via the Stress
System or the Tango+ START/STOP key.
Contact your nearest SunTech Medical
Service Department or authorized service
agent. The unit must be sent to SunTech
Medical for repair.