Herunterladen Inhalt Inhalt Diese Seite drucken

Alarms; Measurement Table; End Test - SunTech Tango+ STRESS BP Gebrauchsanweisung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 87
18

Alarms

Alarms alert you to changes in blood pressure or heart rate. When a limit is
exceeded, an alarm of 5 beeps sound, and a message shows in the status display.
Alarms can be set for systolic blood pressure being too high (SYS HIGH) or too rapid
of a decrease (SYS DROP), and for diastolic blood pressure or heart rate being too
high, DIA HIGH or HR HIGH, respectively.
Sys High:
Sys Drop:
Dia High:
HR High:

Measurement Table

The Measurement Table shows the most recent blood pressure measurements in a
tabular format.
Up or Down:
Select:

End Test

Use this menu to clear readings and prepare for a new patient.
Yes:
No:
When the systolic BP exceeds this setting, the alarm sounds.
Choose from 50 to 280 mmHg (with increments of 10) or set to
Off.
When a drop in systolic BP between measurements exceeds this
setting an alarm sounds. Choose from 10 to 100 mmHg (with
increments of 5) or set to Off.
When the diastolic BP exceeds this setting, an alarm sounds.
Choose from 50 to 150 mmHg (with increments of 10) or set to
Off.
When the heart rate exceeds this setting, an alarm sounds.
Choose from 30 to 200 beats per minute (with increments of 10)
or set to Off.
Scrolls up or down the most
recent 50 readings.
Exits table and gives you
the option to clear all data
from the table.
Choose this to clear readings and prepare the monitor for a new
patient.
Keeps the readings and settings.
TIME
SYS DIA
12:00
120
80
11:57
120
80
HR
65
65

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis