Do not lean on the centrifuge.
Do not stay within the clearance envelope longer than necessary for operational reasons.
Do not allow liquids to enter the equipment.
Do not place the equipment upside down on the platform.
Disconnect the equipment from the power supply when cleaning.
Service should only be performed by authorized personnel.
Do not deposit any potentially hazardous materials within the clearance envelope.
Do not unscrew the two halves of the housing.
Only listed OHAUS original rotors and accessories for this instrument shall be used. Any other
use or intended use is considered improper. OHAUS is not liable for damage resulting from
improper use.
During centrifugation, the presences of people are prohibited within 30 cm around the centrifuge.
Under no circumstances open the lid of the centrifuge while the rotor is still running or rotating
with a speed of > 2m/s.
WARNING: Never work in an environment subject to explosion hazards! The housing
of the instrument is not gas tight. (explosion hazard due to spark formation, corrosion
caused by the ingress of gases).
WARNING: Electrical shock hazards exist within the housing. The housing should only
be opened by authorized and qualified personnel. Remove all power connections to the
unit before opening.
WARNING: The centrifuge is not sealed. Use suitable protection measures when using
the centrifuge for infectious and pathogenic samples. Follow appropriate safety
precautions when handling these samples.
Danger and precautions
To protect people and environment the following precautions should be observed:
During centrifugation, the presences of people are prohibited within 30 cm around the centrifuge
according to the regulations of EN 61010-2-020.
The equipment is not explosion-proof and must therefore not be operated in explosion-endangered
areas or locations. Centrifugation of flammable, explosive, radioactive, or such substances, which
chemically react with high energy, is strictly prohibited. The final decision on the risks associated
with the use of such substances is the responsibility of the user of the centrifuge.
Never spin toxic or pathogenic material without adequate safety precautions, i.e. centrifugation of
buckets / tubes with missing or defective hermetic sealing is strictly prohibited. The user is obliged
to perform appropriate disinfection procedures in case dangerous substances have contaminated
the centrifuge and or its accessories. When centrifuging infectious substances, always pay
attention to the general laboratory precautions. If necessary, contact your safety officer!
It is prohibited to run the centrifuge with rotors other than listed for this unit.
Under no circumstances open the lid of the centrifuge while the rotor is still running or rotating with
a speed of > 2m/s.

2.INSTALLATION

2.1 Unpacking

Carefully remove your centrifuge and each of its components from the package. The included
components vary depending on the centrifuge model (see table below). Save the packaging to ensure
TM
Frontier
2000 Series Centrifuge