Herunterladen Diese Seite drucken

neurosign 4203-NM Gebrauchsanweisung Seite 2

Wiederverwendbare kabel für laternenförmige kehlkopfelektroden

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 3
INSTRUCTIONS FOR USE
English
REUSABLE CABLES FOR LANTERN LARYNGEAL ELECTRODE
Note: Avoid contact with strong, aromatic, chlorinated, ketone, ether, or ester solvents.
Prolonged immersion (more than 1 hour) in alcohol or mild solvents, detergent solutions or
highly alkaline solutions may affect the cable and electrode jacket flexibility.
Disinfection
After cleaning, the reusable cable for Lantern Laryngeal Electrodes may be disinfected with
the use of 70% ethanol. The cable should be wiped with a soft cloth dipped in a 70% ethanol
for at least 1 minute, to ensure proper disinfection of the product.
Drying
After cleaning and disinfection, dry the cable with a clean, soft, low-linting cloth to remove any
remaining water or disinfection solution.
Sterilization
Reusable cables for Lantern Laryngeal Electrodes are not designed to sustain any type of
sterilization, as it may damage the product.
Inspection
Visually inspect the cable for deterioration or excessive wear prior to each use. If cable or
connectors have any fracture, damaged wire insulation, or unnatural angle, discard the cable
according to your hospital's procedure for medical waste. The integrity of the cable can be
checked with a simple resistance meter.
Storage and Transport
Protect cables from damage and contamination during storage and transport. When not in
use, cables should be loosely coiled and stored at room temperature. Do not wrap cables
around the equipment casing to avoid damaging internal wires. When transporting the
products outside their main area of use, pack them in a resealable plastic bag to prevent
damage or contamination.
Note: All serious incidents associated with the use of reusable cables for Lantern Laryngeal
Electrodes should be reported to Technomed Europe at quality@technomed.nl and to a
competent authority of the country where user or patient is established.