measurement (by taking a sample using the skimmer) and compare this value with the
display on the dosing system.
[?] My manual measurement for pH deviates considerably from the display.
➔ Check the pH – sensor (see calibration instructions in the operating instructions)
____________________________________________________________________
[?] After calibration, the error message appears: [Display: Error zero point,
Error slope or Calibriation error]
Possible causes:
➢ Sensor defective
➢ Incorrect buffer solution used
➢ Buffer solutions too old (more than 1 year)
➔ Use a new sensor or buffer solution
Other potential causes:
➢ Defective connection between the sensor and board, e.g. defective measuring
cable of sensor or connector for sensor/device (SN6 or BNC)
➢ Defective measuring cable from connection to board or electronics
Short circuit ➔ Fixed display around value pH 7, since 0 mV = pH 7
Broken cable ➔ Open channel, i.e. display fluctuates or shows "- -"
The measuring duct can be checked with a simulator (special accessory).
____________________________________________________________________
[?] Calibration was successful. Nevertheless, the value still deviates
considerably from the manual measurement.
In this case, a water sample (close to skimmer) must be taken. Then dip the pH sensor
in the buffers for pH 4, pH 7 and the water sample.
pH 7
All values agree. Device is fine. Possible causes:
➢ Incorrect manual measurement
➢ Outdated buffer solution
pH 4
Pool sample
50
Measuring cell