performance is reduced proportionally. Thus, for a deviation of e.g. Δ 50 mV, the pump
only operates with 50 % of the maximum performance.

3.8 Rx limit [120 min]

The dosing time limit prevents a dangerous incorrect dosage if the electrode fails. The
value must be adjusted in line with the pool size. Example: In the event of an accident,
with a set dosing time limit of 60 minutes, a maximum of 1.6 litres of water care product
can be dosed (pump performance 1.6 l/h). After this, the system switches off
automatically. Limit appears in the display (Rx); the triggered alarm can be
acknowledged by pressing the MODE button.
4

Switch-on delay

Corresponds to the start delay of the control unit in minutes. After each start of the
System, the dosing is locked for 8 minutes. By pressing the MODE key, it is possible to
cancel the switch-on delay.
5

Test and special functions

This area is to be operated by authorised personnel only.
MODE
Pump pH: OFF
Pump Rx: OFF
Level pH: U
Level Rx: U
Input pH:
7.2
Input Rx:
650mV
Rx: Contr. Pulse
length
Pressing the MODE key for a longer time (8 seconds)
takes you to the factory settings. Press briefly to
change the sub-items. By pressing
for 2 seconds exits the factory settings.
You can change the values using the arrow keys.
Selection ON/OFF (pump test)
Checking the respective suction lance (option)
Display of a full or empty container at
Connection of a suction lance
Display of the currently measured values at
The respective input
Setting to pulse lengths or to 2-point control