Herunterladen Inhalt Inhalt Diese Seite drucken

Language; Cleaning And Disinfection - Sunoptic Surgical Titan X450 Gebrauchsanleitung

Xenon-lichtquelle mit digitaler dimmfunktion
Vorschau ausblenden Andere Handbücher für Titan X450:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 20

6.2.2.4.5 Language

The language sub menu allows users to change the display language. Use the
to confirm the selection.
NOTE: To reset language back to English, press the
When the icon

7. CLEANING AND DISINFECTION

WARNING: Ensure that the illuminator is de-energized and disconnected from mains power before
attempting to clean and disinfect.
The illuminator can be wiped down with commercially available cleansers commonly used for
disinfection of electronic equipment in hospitals such as ethyl or isopropyl alcohols, disinfecting sprays
containing quaternary ammonium compounds, or hydrogen peroxide.
WARNING: DO NOT use strongly caustic or acidic cleansers such as "Clorox" hypochlorite bleach,
ammonia, muriatic acid, or similar products. DO NOT use acetone, methyl ethyl ketone, or halogenated
/ chlorinated hydrocarbon solvents or cleansers containing any of these restricted compounds.
Apply cleaning agents by light spray or dampened towels. Do not pour liquids onto the device. Do not
allow liquids to enter the device seams or ventilation openings.
Follow all applicable bloodborne pathogen procedures as required by OSHA and/or your hospital
when cleaning and disinfecting the product.
WARNING: The illuminator is not sterilizable. DO NOT attempt to autoclave the devic
LIT-219 Sunoptic Surgical
Rev. F
and
icons to cycle through the options and press the
appears, press it.
®
Page 16 of 147
icon to cycle through the menu.
icon
(English)

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis