Herunterladen Inhalt Inhalt Diese Seite drucken

Sunoptic Surgical Titan X450 Gebrauchsanleitung Seite 11

Xenon-lichtquelle mit digitaler dimmfunktion
Vorschau ausblenden Andere Handbücher für Titan X450:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 20
6.2.1.3 Incompatible Lamp Module Installed
6.2.1.4
Incorrect Wattage Lamp Installed
6.2.1.5
Lamp has Exceeded Usable Life
A notice is displayed in the topleft corner of the home screen when a lamp module has reached 950
hours of use. This notice will begin to flash when the lamp module exceeds 975 hours of use.
LIT-219 Sunoptic Surgical
Rev. F
ACTION: Install correct SSX0450 lamp module.
The illuminator digital control system communicates with
specific Sunoptic Surgical lamp modules. Only Sunoptic
Surgical lamp modules can be installed in the system.
ACTION: Install correct SSX0450 lamp module.
The currently installed lamp module does not match the
intended power of the device. The illuminator requires
400W lamp.
Occurs when lamp module has been used more than 1000
hours.
ACTION: Install a new SSX0450 lamp module or one with
less than 1000 hours of runtime.
®
Page 11 of 147
NOTE: The main screen always displays the lamp
module runtime hours at the top of the screen.
(English)

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis