Herunterladen Inhalt Inhalt Diese Seite drucken

Gima DOMINO 32803 Gebrauchsanweisung Seite 15

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 27
15
3. The pump stop inflating and pressure begins to decrease gradually, during which the user's blood
pressure and pulse will be calculated as Fig.11;
4. There will be a long beep following the accomplishment of measurement. The air in the cuff will deflate
quickly and the blood pressure reading, pulse reading will show in the display. Moreover the meas-
uring time will also display together in two screens alternately. At the same time, the '
remind the user to record the reading as Fig.12;
5. Press button 'M1' or button 'M2' to record the reading in corresponding memory. For example, if
button 'M2' is pressed, the display will show as Fig.13. If the user does not press button, the reading
won't be recorded;
6. Press the button 'START' to return to standby mode. Please rest for at least 3 minutes for another
measurement. The device keeps unused for 3 minutes, the device will be return to standby mode
automatically.
Automatic inflation
There are 4 given levels of given inflation pressure for this device: 190mmHg, 230mmHg, 270mmHg
and 300mmHg.
When 190mmHg is not enough or movement of arm occurs, the device will automatically inflate to rea-
sonable pressure level to ensure a successful measurement. It is not a fault.
Rapid deflation during measurement
If you do not feel well during measurement or want to stop the measurement for some reason, you can
press the START button. The device will quickly release the air in cuff and the device will be returned to
standby mode.
The indicator displays a segment, based on the current data, corresponding to the WHO classification.
For example, if your blood press is 145mmHg (Systolic Pressure), 88mmHg (Diastolic Pressure), ac-
cording to the world health organization standard, your blood pressure level is Mild Hypertension
Note: If the systolic blood pressure and diastolic blood pressure fall into different categories, the higher
value should be taken for classification.
FUNCTION OF MEMORY
Memory recall
1. Domino can store 60 sets of readings each in ' ' and ' ' , and will automatically calculate the average
value of the latest 3 readings for 'M1' and 'M2' respectively. When the memory is full (60 sets of read-
ings are stored), the oldest reading will be replaced by new one. Memory will not clear away even if
power supply is removed;
2. After a measurement is finished or when the device stands by, the user can press button 'M1' or but-
ton 'M2' to recall memory. Press button 'M1' or 'M2', the display will show the average value of the
latest 3 readings as Fig.14;
ENGLISH
' will flash to

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Domino 32804

Inhaltsverzeichnis