Guidance and Manufacturer's Declaration

Per IEC 60601-1-2

electromagnetic emissions

The Naída CI is intended for use in the electromagnetic environment specified below. The customer or
the user of the Naída CI should ensure that it is used in such an environment.
emissions Test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/ flicker emissions
IEC 61000-3-3

electromagnetic Immunity

The Naída CI is intended for use in the electromagnetic environment specified below. The customer or
the user of the Naída CI should ensure that it is used in such an environment.
IeC 60601
Immunity Test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
Radiated RF
80 MHz to
IEC 61000-4-3
58
Naída CI Sound Processor User Guide
Compliance
Not applicable
Not applicable
Compliance level
Test level
± 6 kV
± 6 kV contact
contact
± 8 kV air
± 8 kV air
3 A/m
3 A/m
3 V/m
3 V/m
2.5 GHz
electromagnetic environment-Guidance
The Naída CI uses RF energy only for its
internal function. Therefore, its RF emissions
Group 1
are very low and are not likely to cause any
interference in nearby electronic equipment.
Class B
establishments, including domestic establishments
and those directly connected to the public
low-voltage power supply network that supplies
electromagnetic environment-Guidance
a
Floors should be wood, concrete or ceramic tile. If
floors are covered with synthetic material, the relative
Power frequency magnetic fields should be at
levels characteristic of a typical location in a typical
commercial or hospital environment.
Portable and mobile RF communications equipment
should be used no closer to any part of the Naída CI,
including cables, than the recommended separation
distance calculated from the equation applicable to
Recommended separation distance
P
where
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in metres (m). Field strengths from fixed RF
transmitters, as determined by an electromagnetic
site survey,
Interference may occur in the vicinity of
equipment marked with the following symbol:
The Naída CI is suitable for use in all
buildings used for domestic purposes.
humidity should be at least 30%.
the frequency of the transmitter.
d
= 1.2√P < 800 MHz
d
= 2.3√P ≥ 800 MHz
is the maximum output power rating of the