Charger LeD (Light emitting Diode) Indicator

LeD Port Lights
LeD Micro USB Light
The charger has different indicator lights that are located at the base of each PowerCel charging
port as well as where the power cable enters the charger. If the green micro USB light is lit, it
means the charger is connected to a power source and ready to charge the PowerCels.
The LEDs located at the base of each PowerCel charging port have 3 color options.
Understanding PowerCel Charging Port LEDs
Indication
Color
PowerCel is charging
Blue
Green
PowerCel is fully charged
Red
Fault Condition
If a fault is exhibited by the PowerCel charger, find out if the fault is with the PowerCel or the
charger itself by doing the following. Insert a different PowerCel in the charger bay that exhibits
the problem. If the red LED is still on, there is a problem with the charger. If the red LED does
not light up, there is a problem with the PowerCel. If there is a problem with either the PowerCel
or the charger, please contact Advanced Bionics.
The charger LED turns off (or may show a faint blue if the room is very dark to indicate either
the battery is fully charged or there is a fault with the battery. The user should wait one full
minute. If the LED turns green, the battery is fully charged and ready for use. If the LED does not
turn green after one minute, the battery should be considered faulty and you should contact
Advanced Bionics for a replacement.