Herunterladen Inhalt Inhalt Diese Seite drucken

Agilent Technologies TwisTorr 74FS AG Rack Bedienungshandbuch Seite 236

Inhaltsverzeichnis

Werbung

17
Technical Information
Letter Protocol Description
When the Request is "J" (Numerical readings), the Answer will
contain a string of 5 characters with the following parameters:
Tab. 12
BYTES
1
2
3
4
5
When the Request is "K" (Counters readings), the Answer will
contain a string of 11 characters with the following parameters:
Tab. 13
BYTES
1-4
5-8
9-10
11
On request a sample program in QBasic language is available by
Agilent.
236/300
MEANING
Current (0-255 scaled)
Voltage (0-255 scaled)
Rotational speed KRPM
Pump temperature °C (0-254 temperature reading, 255 =
fail)
CRC
MEANING
Cycle time
Pump life
Cycle number
CRC
TwisTorr 74 FS AG Rack Controller User Manual / 87-901-055-01