Herunterladen Inhalt Inhalt Diese Seite drucken

Agilent Technologies TwisTorr 74FS AG Rack Bedienungshandbuch Seite 206

Inhaltsverzeichnis

Werbung

16
Instructions for Use
Error Messages
MESSAGE
TOO HIGH
LOAD
SHORT
CIRCUIT
SYSTEM
OVERRIDE
OVERVOLTAGE
POWER FAIL
206/300
DESCRIPTION
The current drawn by the pump is
too high.
A short circuit is present on the
pump connector.
The pump is stopped by an
emergency stop signal provided via
a remote contact.
Controller power supply circuitry is
faulty or the Controller received
an unexpected excessive voltage.
Failure in the power supply section.
TwisTorr 74 FS AG Rack Controller User Manual / 87-901-055-01
REPAIR ACTION
Check that the pump rotor is
free to rotate. Press the
START push-button twice to
start the pump.
Check connections and short
circuits between pump and
controller. Press the START
push-button twice to start the
pump.
Remove the controller power
cable and check the
emergency condition. Then
reconnect the power cable
and press the START push-
button twice to start the
pump.
Press the START push-button
twice to start the pump.
Should the message still be
present, call the Agilent
service.
Contact Agilent for
Maintenance.

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis