3.6.4 Trouble shooting
Do not attempt to repair on your own. Warranty will be invalid if unauthorized repair has been
carried out. Due to safety reasons repair and maintenance may only be done by the manufacturer.
Please note:
For further technical data and information, please refer to the separate and comprehensive inst-
ructions for use, enclosed with the product, isolating transformer and insulating monitor.
3.7

Cable connection

28. Anyone operating the equipment cart or integral isolating transformers must ensure that there
is a only a detachable connection with the tool for the cable connection between the multi-socket
outlets on the cart and the equipment. Our line of accessories includes a selection of covers for
multiple power outlet strips.
3.8

Combination of equipment

The following should be observed for combination of equipment on the carrier system:
Auxiliary equipment connected to analogue and digital interfaces of the equipment must
be certified in compliance with the relevant EN specifications (e. g. IEC 60950 for data-
processing equipment and IEC 60601-1 for medical electrical equipment).
Furthermore, all configurations must comply with the valid version of the standard
IEC 60601-1. Anyone connecting additional equipment to the signal input or signal
output is a system configurer and is therefore responsible for ensuring compliance with the
valid version of the standard IEC 60601-1.
If you have any questions, you should contact your local dealer or technical services.
Please note:
Make sure that this is also the case for adaptation of equipment in the power supply circuit (e. g.
multiple socket strip).
3.9
EMC
Electromagnetic compatibility of the medical electrical equipment positioned on the cart has to
be checked by the overall system configurer. If you use a different equipment combination, you
should check electromagnetic compatibility between individual items of equipment before using in
a medical setting.
Customer-specific equipment trollies used within the nuclear spin environment must be tested
by the customer for suitability for use due to the ferromagnetic materials they contain. ITD GmbH
excludes any liability in this respect.
Operating manual for mobile equipment carriers 09/2020
ELECTRICAL SAFETY
41