If the pressure sensor is not properly calibrated after the autoclave has been installed, the autoclave 
may refuse to start a program. 
 
7.5 Validation and calibration 
If you have any questions about validation and calibration of the instrument, please contact the manufacturer. 
8 Sterilization programs 
Program 
Vacuum 
phases 
121 °C / SOLID 
121 °C / POROUS 
121 °C / HOLLOW 
134 °C / SOLID 
134 °C / POROUS 
134 °C / HOLLOW 
B‐D TEST 
VACUUM TEST 
CLEAN PROCESS 
PRIONS 
LIQUID 
 
You can create a new custom program or modify an existing program at any time. 
The  SOLID  programs  can  only  sterilize  solids  that  are  not  wrapped.  Such  as  unwrapped  forceps,  scissors, 
tweezers, trays, etc. 
The "Porous" programs are specially designed for bodies made of porous material. Three deaeration steps by 
means  of  the  vacuum  pump  ensure  complete  penetration  of  the  steam  into  the  porous  material.  This  is  a 
necessary condition for the proper sterilization of porous materials. 
The "Hollow" programs can sterilize type A and type B hollow items. These are, for example, glasses or cannulas.  
CertoClav Sterilizer GmbH
Peintner Str. 10 
4060 Leonding, Austria 
www.certoclav.com 
  
support@certoclav.com
Sterilization 
time 
 
[min]
20 
20 
20 
‐ 
‐ 
19 
‐ 
20 
 
Sterilization 
Sterilization 
temperature 
 
pressure 
[°C]
121 
121 
121 
134 
134 
134 
134 
‐ 
105 
134 
121 
Vacuum hold 
 
time
 
[kPa]
 [min]
110 
‐ 
110 
‐ 
110 
‐ 
210 
‐ 
210 
‐ 
210 
‐ 
210 
‐ 
‐80 
15 
20 
‐ 
210 
‐ 
110 
‐ 
 
Seite / Page 78 
Drying 
time 
 
[min]
10 
10 
‐ 
10 
10 