CertoClav_Vakuum_Pro_12‐22_DE_EN_V18.docx
If the pressure sensor is not properly calibrated after the autoclave has been installed, the autoclave
may refuse to start a program.
7.5 Validation and calibration
If you have any questions about validation and calibration of the instrument, please contact the manufacturer.
8 Sterilization programs
Program
Vacuum
phases
121 °C / SOLID
121 °C / POROUS
121 °C / HOLLOW
134 °C / SOLID
134 °C / POROUS
134 °C / HOLLOW
B‐D TEST
VACUUM TEST
CLEAN PROCESS
PRIONS
LIQUID
You can create a new custom program or modify an existing program at any time.
The SOLID programs can only sterilize solids that are not wrapped. Such as unwrapped forceps, scissors,
tweezers, trays, etc.
The "Porous" programs are specially designed for bodies made of porous material. Three deaeration steps by
means of the vacuum pump ensure complete penetration of the steam into the porous material. This is a
necessary condition for the proper sterilization of porous materials.
The "Hollow" programs can sterilize type A and type B hollow items. These are, for example, glasses or cannulas.
CertoClav Sterilizer GmbH
Peintner Str. 10
4060 Leonding, Austria
www.certoclav.com
support@certoclav.com
Sterilization
time
[min]
1
20
3
20
3
20
1
4
3
4
3
4
3
4
‐
‐
3
5
3
19
‐
20
Sterilization
Sterilization
temperature
pressure
[°C]
121
121
121
134
134
134
134
‐
105
134
121
Vacuum hold
time
[kPa]
[min]
110
‐
110
‐
110
‐
210
‐
210
‐
210
‐
210
‐
‐80
15
20
‐
210
‐
110
‐
Seite / Page 78
Drying
time
[min]
3
7
10
3
7
10
7
‐
10
10
0