Herunterladen Inhalt Inhalt Diese Seite drucken

DJO Donjoy IceMan CLEAR3 Bedienungsanleitung Seite 13

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 38
TROUBLESHOOTING
Pump will not turn on
• Check all electrical connections and make sure the wall plug has power.
Cold pad will not cool down
• Make sure trapped air is out of cold pad once power is applied.
• Make sure cooler is filled with ice and water.
• Check all hose connections and make sure cold pad is wrapped properly to
allow water to flow.
• Check and clean filter cap located under the pump assembly. First begin
by draining the cold therapy unit. Then remove the filter cap located
underneath the pump assembly by unsnapping it. Clean the filter cap,
removing any material that may be clogging the filter. Finally, reinstall the
filter cap before using the device.
Cold pad pressure is low
• Check water level. Add water if necessary.
• Make sure trapped air is out of cold pad.
No water flow
• Check water level. Add water if necessary
• Check and clean filter cap located under pump assembly.
• Cold pad and hose are wrapped too tightly or the hose is kinked. Unwrap
and rewrap the cold pad with cold therapy unit running, making sure water is
circulating freely throughout the cold pad and the hose is not kinked.
• Make sure connection between the cold therapy unit hose and cold pad hose
is properly connected.
• With cooler filled with water. Turn off device.  Detach cold pad from the
motorized unit.  Insert a blunt instrument (such as a pen) into each of the
female fluid connectors on the motorized unit tube set to depress and open
the spring loaded shut-off valves.  Depress each valve for 1-2 seconds each to
allow any trapped air to escape.  Reconnect pad and turn device back on.
Water leak at connector
• Connector between cold therapy unit hose and cold pad hose is not properly
connected. Stop machine, disconnect hose, and reconnect hose listening for
a "snap" or "click", and restart the unit.
• Check barrier to ensure it is dry. Check o-rings. Replace with dry barrier, if it
is wet.
Water leak at cold pad
• Faulty cold pad, replace.
• Check barrier to ensure it is dry. Replace with dry barrier, if it is wet.
Magnetic or Electromagnetic Interference
• Reorient or relocate the device. Increase the separation between the
equipment. Plug the power supply into an outlet on a circuit different from
that to which the other device(s) are connected.