System Error
Display: „005"
Solution: Replace the batteries first. If Err
appears again, please contact your local dea-
ler.
Test result is higher than 630 mg/dL (35.0
mmol/L)
Display: „HI"
Solution: Test again. If the result is still the
same, try a control solution test and if the
control solution test falls within the correct
range, please call your healthcare professional
immediately.
Test result is under 20 mg/dL (1.1 mmol/L)
Display: „LO"
Solution: Test again. If the result is still the
same, try a control solution test and if the con-
trol solution test falls within the correct range,
please call your healthcare professional imme-
diately.
Err message. Not enough sample was drawn into
the test strip for measurement.
Display: „Err & Blood Drop"
Solution: Discard the test strip and repeat the
test.