The filter must not be cleaned and reused.
The filter must not be removed for servicing while
the EMSER
Inhalator Pro is in use.
®
Only use genuine filters! Do not use the device
without a filter!
6. Malfunctions and countermeasures
The device will not switch on.
•
Ensure that the power cable is inserted correctly.
•
Ensure that the operating switch ( 1 a) is in position I.
•
Ensure that the unit has been used within the limits
of the operating time specified in these instructions.
The atomiser is working poorly or not working at all
•
Ensure that the compressed air hose ( 2 ) is correctly
connected at both ends.
•
Check that the compressed air hose is not crushed,
kinked, dirty or blocked. Replace it with a new one
if necessary.
•
Check that the atomiser ( 3 ) is fully assembled and
that the coloured nebuliser head ( 3 a) is attached cor-
rectly and not blocked.
•
Ensure that the medication has been added to the
device.
7. Warranty/service
We offer a 5-year product warranty from the date of pur-
chase against material and production faults. The war-
ranty does not cover:
•
Damage due to improper operation.
•
Wearing parts.
•
Cases of personal negligence on the part of the cus-
tomer.
•
Cases in which the device has been opened by a
non-authorised workshop.
The warranty does not affect the customer's statutory
rights. To assert a warranty claim within the warranty
period, the customer must provide proof of purchase.
The warranty claim must be submitted within a period
of 5 years from the date of purchase to Sidroga GfGm-
bH, Bad Ems, for Germany or Austria, or to Sidroga AG,
Rheinfelden, for Switzerland.
In case of complaints, please contact our service depart-
ment using the following contact details:
Germany/Austria:
Service hotline:
Tel.: +49 ( 0 ) 2603/9604-710
Fax: +49 ( 0 ) 2603/9604-711
E-Mail: info@emser.de
www.emser.de
If requested, please send the faulty product to the fol-
lowing address:
Sidroga Gesellschaft für Gesundheitsprodukte mbH
Postfach 1262
DE-56119 Bad Ems
GERMANY
Switzerland:
Customer service:
Tel.: +41 61 875 00 75
Fax: +41 61 875 00 78
E-Mail: apothekenservice@sidroga-pharma.com
www.emser.ch
If requested, please send the faulty product to the fol-
lowing address:
Sidroga AG
Weidenweg 15
CH-4310 Rheinfelden
SWITZERLAND
8. Technical data
Type:
AEROSOL PERFORMANCES
ACCORDING TO
EN27427:2019 based on
adult ventilatory pattern
with Sodium Fluoride (NaF): 1 ) Aerosol output: 0.11 ml
24
GCE843
2 ) Aerosol output rate: 0.07 ml/
min
3 ) Percentage of fill volume
emitted per min: 3.5 %
4 ) Residual Volume: 1.68 ml
5 ) Particle size (MMAD): 3.07 µm
6 ) GSD (geometric standard de-
viation): 2.03
7 ) RF (respirable fraction < 5 µm):
75 %
8 ) Large particle range (> 5 µm):
25 %