6.5 Status and error codes; datalogger type ...P4
6.5.1 Datalogger display
ALA
Additional alarm text in power-save mode
Strt
The datalogger is waiting for log start in accordance with the settings at menu item:
Datalogger setup
StOP
The logger is in Start/Stop mode and the memory is full.
No further measured values can be logged. Reprogram the logger to restart logging.
This is the delivery status of the PT100; datalogger type ...P4.
b.F.
Battery voltage is too low
C.F.
Faulty datalogger
HHHH
Faulty datalogger keypad
L.C.
Faulty datalogger
O.F.
Measured value exceeds allowable maximum
U.F.
Measured value is lower than allowable minimum
6.5.2 Status of datalogger in elproLOG ANALYZE ...P4
- System resets since last reprogramming
- This error message appears at datalogger status in line: „Last reprogrammed". It is the result of
the datalogger reset counter.
- All entries in the alarm protocol will be erased!
6.6 Notes; datalogger type ...P4
Software
ECOLOG-NET dataloggers require evaluation software, version 3.31 or higher.
No Connection (USB or LAN)
Reaction: enlarge the timeout parameters in communication settings:
- elproLOG ANALYZE (Options - Com Port... - Extended)
- elproLOG MONITOR (Tools - Options - Communication Timeouts)
CD-ROM
Refer the elproLOG ANALYZE CD-ROM for further product information.
6.7 Measuring range, operating range, accuracy and supply; datalogger type ...P4
Temperature measurement
(Logger only;
logger at room temperature)
Additional error caused by cable asymmetries for length over 50m
Additional error caused by 2 or 3 wire PT100 sensor
Operating range
Protective class
Supply
Seite - 32
Manual
ECOLOG-NET
Range
Resolution
-200.0°C..-100.0°C
0.1°C
-99.9°C..400.0°C
0.1°C
400.1°C..500.0°C
0.1°C
500.1°C..600.0°C
0.1°C
length
type
50m
4x0.25mm
100m
4x0.34mm
Wire
Cable
2
2x0.366mm
3
3x0.366mm
LP4: -40°C ... 55°C
WP4: -20°C ... 55°C
IP30
U
: 10 ...30VDC LP4: 12VDC; 140mA
B
WP4: 12VDC; 200mA
Linearity
± 0.3°C
± 0.2°C
± 0.3°C
± 0.5°C
typical Error
2
, sheelded
± 0.1°C
2
, sheelded
± 0.2°C
typical Error
2
, sheelded
± 0.27°C/m
2
, sheelded
± 0.21°C/m
24VDC; 70mA
24VDC; 100mA
D-EL-6003Bk
11. 2009 AG

Werbung

loading

Diese Anleitung auch für:

Ecolog-net lh2Ecolog-net wh2Ecolog-net wp4